L&D Surgical Lights W81K0026QA200
Summary
AI-generated · Jun 13, 2026The lighting system must allow clinicians to adjust the focal point and working distance directly from the center sterile handle without moving the light head or leaving the sterile field. Vertically segmented reflectors must capture upward-directed light and redirect it downward to create a crisp, homogeneous spot with minimized shadows. It must provide dual selectable color temperatures (4100K soft white and 4500K bright white) adjustable for specific procedures, deliver up to 160,000 lux, and offer light head diameter options of 24 or 30 inches. The architecture must support a manual focus mode using high‑intensity LEDs with a centrally positioned ring of fixed‑focus LEDs and be explicitly compatible with flat-screen displays, booms, and HD surgical cameras.
The suspension and arm system must provide at least 110 degrees of vertical travel for steep downward positioning, and all axes of the lighting fixture must rotate 360 degrees continuously without hard spots. The optical design should allow the fixtures to be placed farther from the patient while maintaining precise detail.
Clinical Salient Requirements: ?System must allow clinicians to adjust the focal point and working distance directlyfrom the center sterile handle without moving the light head or leaving the sterilefield. ?The light head must utilize vertically segmented reflectors to capture upward-directed light and redirect it downward, creating a crisp, homogenous spot thatminimizes shadows. ?The system must offer dual selectable color temperatures (specifically 4100K softwhite and 4500K bright white) adjustable for specific procedure types. Technological Salient Requirements: Server Requirements: ?The lights must deliver a maximum intensity of 160,000 lux and offer light headdiameter options of 24 inches or 30 inches to fill the surgical cavity with seamlesslight. ?The lighting architecture must integrate a manual focus mode of high-intensity LEDscombined with a centrally positioned ring of fixed-focus LEDs. ?The lights must be explicitly compatible with and ready for integration with flat-screen displays, booms, and HD surgical cameras. System Requirement: ?The suspension and arm system must provide at least 110 degrees of vertical travelto allow steep downward positioning. ?All axes of the lighting fixture must feature 360-degree continuous rotation withouthard spots. ?The optical design must allow the fixtures to be placed further away from the patientwhile maintaining precise detail.
From Solicitation posted on Jun 12, 2026Clinical Salient Requirements: ?System must allow clinicians to adjust the focal point and working distance directlyfrom the center sterile handle without moving the light head or leaving the sterilefield. ?The light head must utilize vertically segmented reflectors to capture upward-directed light and redirect it downward, creating a crisp, homogenous spot thatminimizes shadows. ?The system must offer dual selectable color temperatures (specifically 4100K softwhite and 4500K bright white) adjustable for specific procedure types. Technological Salient Requirements: Server Requirements: ?The lights must deliver a maximum intensity of 160,000 lux and offer light headdiameter options of 24 inches or 30 inches to fill the surgical cavity with seamlesslight. ?The lighting architecture must integrate a manual focus mode of high-intensity LEDscombined with a centrally positioned ring of fixed-focus LEDs. ?The lights must be explicitly compatible with and ready for integration with flat-screen displays, booms, and HD surgical cameras. System Requirement: ?The suspension and arm system must provide at least 110 degrees of vertical travelto allow steep downward positioning. ?All axes of the lighting fixture must feature 360-degree continuous rotation withouthard spots. ?The optical design must allow the fixtures to be placed further away from the patientwhile maintaining precise detail.
From Solicitation posted on Jun 16, 2026Notice history
2-
Solicitation Posted Jun 12, 2026
-
Solicitation LATEST Posted Jun 16, 2026No changes from previous notice
Details
Award Information
Not yet awarded
Contacts
Agency
Place of Performance
USA