L&D Surgical Lights W81K0026QA200

Solicitation W81K0026QA200 Copied Notice ID f5b1b23993204a9a99b7ad6adfd7c097 Copied DEPT OF DEFENSE — W40M MRC0 WEST
SAM.gov
Posted
Jun 16, 2026
Deadline
Jun 22, 2026
Set-aside
SDVOSBS
NAICS
339113
PSC
6530

Summary

AI-generated · Jun 13, 2026

The lighting system must allow clinicians to adjust the focal point and working distance directly from the center sterile handle without moving the light head or leaving the sterile field. Vertically segmented reflectors must capture upward-directed light and redirect it downward to create a crisp, homogeneous spot with minimized shadows. It must provide dual selectable color temperatures (4100K soft white and 4500K bright white) adjustable for specific procedures, deliver up to 160,000 lux, and offer light head diameter options of 24 or 30 inches. The architecture must support a manual focus mode using high‑intensity LEDs with a centrally positioned ring of fixed‑focus LEDs and be explicitly compatible with flat-screen displays, booms, and HD surgical cameras.

The suspension and arm system must provide at least 110 degrees of vertical travel for steep downward positioning, and all axes of the lighting fixture must rotate 360 degrees continuously without hard spots. The optical design should allow the fixtures to be placed farther from the patient while maintaining precise detail.

Clinical Salient Requirements: ?System must allow clinicians to adjust the focal point and working distance directlyfrom the center sterile handle without moving the light head or leaving the sterilefield. ?The light head must utilize vertically segmented reflectors to capture upward-directed light and redirect it downward, creating a crisp, homogenous spot thatminimizes shadows. ?The system must offer dual selectable color temperatures (specifically 4100K softwhite and 4500K bright white) adjustable for specific procedure types. Technological Salient Requirements: Server Requirements: ?The lights must deliver a maximum intensity of 160,000 lux and offer light headdiameter options of 24 inches or 30 inches to fill the surgical cavity with seamlesslight. ?The lighting architecture must integrate a manual focus mode of high-intensity LEDscombined with a centrally positioned ring of fixed-focus LEDs. ?The lights must be explicitly compatible with and ready for integration with flat-screen displays, booms, and HD surgical cameras. System Requirement: ?The suspension and arm system must provide at least 110 degrees of vertical travelto allow steep downward positioning. ?All axes of the lighting fixture must feature 360-degree continuous rotation withouthard spots. ?The optical design must allow the fixtures to be placed further away from the patientwhile maintaining precise detail.

From Solicitation posted on Jun 12, 2026

Clinical Salient Requirements: ?System must allow clinicians to adjust the focal point and working distance directlyfrom the center sterile handle without moving the light head or leaving the sterilefield. ?The light head must utilize vertically segmented reflectors to capture upward-directed light and redirect it downward, creating a crisp, homogenous spot thatminimizes shadows. ?The system must offer dual selectable color temperatures (specifically 4100K softwhite and 4500K bright white) adjustable for specific procedure types. Technological Salient Requirements: Server Requirements: ?The lights must deliver a maximum intensity of 160,000 lux and offer light headdiameter options of 24 inches or 30 inches to fill the surgical cavity with seamlesslight. ?The lighting architecture must integrate a manual focus mode of high-intensity LEDscombined with a centrally positioned ring of fixed-focus LEDs. ?The lights must be explicitly compatible with and ready for integration with flat-screen displays, booms, and HD surgical cameras. System Requirement: ?The suspension and arm system must provide at least 110 degrees of vertical travelto allow steep downward positioning. ?All axes of the lighting fixture must feature 360-degree continuous rotation withouthard spots. ?The optical design must allow the fixtures to be placed further away from the patientwhile maintaining precise detail.

From Solicitation posted on Jun 16, 2026

Notice history

2
  1. Solicitation Posted Jun 12, 2026 View
  2. Solicitation LATEST Posted Jun 16, 2026
    No changes from previous notice

Details

Solicitation number W81K0026QA200
Notice ID f5b1b23993204a9a99b7ad6adfd7c097
Notice type Solicitation
Product / Service (PSC) 6530
NAICS 339113
Set-aside SDVOSB Sole Source
Place of performance Colorado Springs, Colorado
Archive date Jul 07, 2026

Award Information

Not yet awarded

Contacts

primary
Joseph Flores

Email

Agency

DEPT OF DEFENSE
DEPT OF THE ARMY
US ARMY MEDICAL COMMAND
US ARMY HEALTH CONTRACTING ACTIVITY
W40M MRC0 WEST

Place of Performance

Colorado Springs, Colorado 80913
USA

Dates

Posted Jun 16, 2026 1 month ago
Last Updated Aug 06, 2026 1 day ago
Due Jun 22, 2026 1 month ago