VPRO Max2 Low Temp Sterilizers 36C25526Q0154

Solicitation 36C25526Q0154 Copied Notice ID 4211ea88674841a19b166c31a5acd9f8 Copied VETERANS AFFAIRS, DEPARTMENT OF — 255-NETWORK CONTRACT OFFICE 15 (36C255)
SAM.gov
Posted
Jan 14, 2026
Deadline
Jan 21, 2026
Set-aside
NONE
NAICS
334510
PSC
6515

Summary

AI-generated · Jan 15, 2026

Provide at least three VPRO MaX2 low-temperature vaporized hydrogen peroxide sterilizers that meet the following specs: each unit has cabinets sized about 33 x 40-15/16 x 72 inches or 33 x 38-5/16 x 72 inches for the recessed cabinet, mounted as a single-door cabinet installation, with no plumbing or ventilation required beyond a suitable room ventilation; a dedicated electrical connection is needed and a power cord is supplied. The installation space is approximately 35 by 110 (facility footprint). Sterilizers use vaporized hydrogen peroxide suitable for heat-sensitive materials, maintaining load temperatures not exceeding 55°C (131°F); FDA-cleared devices (with or without plasma) are acceptable.

Per ANSI/AAMI ST58:2013 (or newer), the installation area must provide ventilation of at least 10 air exchanges per hour. Electrical requirements include support for 16 A current, minimum 20 A circuit breaker, NEMA L15-20R plug, and grouped or ganged circuit protection with a disconnect as required by NEC and local codes, plus individual power shutoffs near each unit for servicing.

2. SPECIFICATIONS . Minimum of three low temperature sterilizers shall have a minimum individual dimensions of Cabinet (DD): 33 x 40-15/16 x 72 (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38- 5/16 x 72 (838 x 973 x 1829 mm). o Mounting single door, cabinet o System installation requires no plumbing, ventilation or air o supply only a dedicated electrical connection is needed. A power cord is supplied for this connection o The facility space in which the vehicle can occupy is approx. 35 x 110 . . Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive materials because temperatures within the load currently do not exceed 55C (131F) and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized hydrogen peroxide sterilizers, either with or without plasma, are currently available in the United States. . Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area with a ventilation system capable of at least 10 air exchanges per hour. . VPRO MaX2 must support utility requirements: o Current 16 AMP o Minimum Circuit Breaker 20 AMPS o Use NEMA L15-20R Plug o Provide grouped or ganged circuit protection and disconnect for phase 3 power as required by NEC and local codes. o Individual power shutoffs near each machine for servicing.

From Sources Sought posted on Jan 14, 2026

Notice history

1
  1. Sources Sought LATEST Posted Jan 14, 2026 View

Details

Solicitation number 36C25526Q0154
Notice ID 4211ea88674841a19b166c31a5acd9f8
Notice type Sources Sought
Product / Service (PSC) 6515
NAICS 334510
Set-aside No Set aside used
Place of performance Kansas City, Missouri
Archive date Feb 05, 2026

Award Information

Not yet awarded

Documents

1
View on SAM.gov

Contacts

primary
Jon Lutmer

Email

Phone

Agency

VETERANS AFFAIRS, DEPARTMENT OF
VETERANS AFFAIRS, DEPARTMENT OF
255-NETWORK CONTRACT OFFICE 15 (36C255)

Place of Performance

Kansas City, Missouri 64128
USA

Dates

Posted Jan 14, 2026 6 months ago
Last Updated Aug 06, 2026 1 day ago
Due Jan 21, 2026 6 months ago