VPRO Max2 Low Temp Sterilizers 36C25526Q0154
Summary
AI-generated · Jan 15, 2026Provide at least three VPRO MaX2 low-temperature vaporized hydrogen peroxide sterilizers that meet the following specs: each unit has cabinets sized about 33 x 40-15/16 x 72 inches or 33 x 38-5/16 x 72 inches for the recessed cabinet, mounted as a single-door cabinet installation, with no plumbing or ventilation required beyond a suitable room ventilation; a dedicated electrical connection is needed and a power cord is supplied. The installation space is approximately 35 by 110 (facility footprint). Sterilizers use vaporized hydrogen peroxide suitable for heat-sensitive materials, maintaining load temperatures not exceeding 55°C (131°F); FDA-cleared devices (with or without plasma) are acceptable.
Per ANSI/AAMI ST58:2013 (or newer), the installation area must provide ventilation of at least 10 air exchanges per hour. Electrical requirements include support for 16 A current, minimum 20 A circuit breaker, NEMA L15-20R plug, and grouped or ganged circuit protection with a disconnect as required by NEC and local codes, plus individual power shutoffs near each unit for servicing.
2. SPECIFICATIONS . Minimum of three low temperature sterilizers shall have a minimum individual dimensions of Cabinet (DD): 33 x 40-15/16 x 72 (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38- 5/16 x 72 (838 x 973 x 1829 mm). o Mounting single door, cabinet o System installation requires no plumbing, ventilation or air o supply only a dedicated electrical connection is needed. A power cord is supplied for this connection o The facility space in which the vehicle can occupy is approx. 35 x 110 . . Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive materials because temperatures within the load currently do not exceed 55C (131F) and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized hydrogen peroxide sterilizers, either with or without plasma, are currently available in the United States. . Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area with a ventilation system capable of at least 10 air exchanges per hour. . VPRO MaX2 must support utility requirements: o Current 16 AMP o Minimum Circuit Breaker 20 AMPS o Use NEMA L15-20R Plug o Provide grouped or ganged circuit protection and disconnect for phase 3 power as required by NEC and local codes. o Individual power shutoffs near each machine for servicing.
From Sources Sought posted on Jan 14, 2026Notice history
1Details
Award Information
Not yet awarded
Contacts
Phone
Agency
Place of Performance
USA