Contacts, documents and full notice history are available with a subscription.
Solicitation SDVOSBC Expired 1 notice 1 document

AMSCO Pass Through Dry Cabinet 36C25526Q0323

Solicitation 36C25526Q0323 Copied Notice ID 45136410c38644f6b725679810f96c3d Copied VETERANS AFFAIRS, DEPARTMENT OF — 255-NETWORK CONTRACT OFFICE 15 (36C255)
SAM.gov
Posted
Apr 16, 2026
Deadline
Apr 28, 2026
Set-aside
SDVOSBC
NAICS
334510
PSC
6515

Summary

AI-generated · Apr 17, 2026

Provide and install two pass-through instrument drying cabinets at the Kansas City VA Medical Center to support the Sterile Processing Service in reprocessing reusable medical devices used for invasive procedures, surgeries, and related clinical areas (dental and surgical clinics). Include all required services, labor, travel, materials, shipping, parts, connections, and equipment for on-site installation, plus training for clinical staff prior to use covering all shifts, and removal of debris and packing materials after installation.

Cabinets must meet the following specs and requirements: maximum 78 inches (H) x 38 inches (W) x 31 inches (D) per unit (total space 80 inches H x 76 inches W x 32 inches D); maintain full separation between decontamination and high-level disinfection environments per AAMI ST91-21; support drying of cannulated internal channels and DaVinci robotic arms; built-in HEPA filtration with lumen purge connectors; capable of drying heat‑sensitive devices; maintain 63–77°F (17–25°C) and 30–60% RH; electrical requirements 120Vac, 60Hz, 20A; max 750 lb per cabinet. Site visits to verify conditions and OEM conformance required; only fully qualified technicians allowed. Post-installation, provide a two-year service maintenance agreement with emergency response within 4 hours, temporary solution within 24 hours if needed, and a permanent solution planned within 5 business days; comply with OSHA, NFPA, and VA Design Guide requirements. Documentation of maintenance must be provided in hard copy and electronic form; hours of operation are 7:30 am–4:00 pm, Monday–Friday or as directed by the COR; security requirements specify no Security Accreditation Package and prohibit photography unless authorized.

1. STATEMENT OF WORK The contractor shall provide all services, labor, travel, transportation, material, shipping, parts, connections, and equipment necessary for provision and on-site installation of equipment to provide a minimum quantity of two pass-through instrument drying cabinet units located at 4801 Linwood Blvd, Kansas City MO 64128. The pass-through instrument drying cabinets are needed to support Sterile Processing Service (SPS) for reprocessing reusable medical devices (RMD) used in invasive procedures. They are needed to support medical procedures and surgeries in the operating room, dental clinic, and surgical clinic. The contractor shall also provide training to clinical staff prior to equipment use, capturing all tours of duty for the sterile processing department. The contractor is responsible for removing and disposing of all debris and packing material left from the installation of the pass-through instrument drying cabinets. The contract includes two pass-through instrument drying cabinets, and full installation services to include necessary installation components to support wall structure for pass-through capabilities per schedule below: 2. SPECIFICATIONS Minimum of two (2) pass-through instrument drying cabinets shall have maximum individual dimensions of 78 inches (H) x 38 inches (W) x 31 inches (D), for a total space occupancy of 80 inches (H) x 76 inches (W) x 32 inches (D). The pass-through instrument drying cabinets shall maintain full separation between the decontamination environment and high level disinfection environment with no risk of cross contamination between the areas (AAMI ST91-21). The facility space in which the AER can occupy is approximately 8.2 (L) x 8.8 (h) Ability to support drying of canulated instrumentation s internal channels. Ability to support drying of internal channels of DaVinci robotic arms. Built in HEPA filtrated air with HEPA air purge connectors for lumens. Ability to dry heat-sensitive devices. Ability to maintain temperature between 63F - 77F (17C - 25C) and humidity between 30%-60% RH Instrument drying cabinet must support utility requirements: 120Vac, 60Hz, 20 Amp, 53dBa Maximum weight per cabinet of 750 lb (340kg) 3. REQUIREMENTS It is the responsibility of the Contractor to conduct site visits, verify existing conditions, quantities of materials, installation instruction of materials to be used, and discrepancies or conflicting requirements in the Statement of Work before preparing the Final Proposal. The contractor shall provide and install two pass-through instrument drying cabinets to the VA Medical Center located at 4801 Linwood Kansas City, Mo 64128. The contractor shall ensure that all repairs, preventable, and remedial maintenance provided during the provision timeframe of this contract meet and or exceed the OEM s specifications. The contractor shall ensure that only FULLY QUALIFIED TECHNICIANS/ENGINEERS are employed in the performance of all work performed under this contract. The contractor shall ensure that the highest standard of professional capability and workmanship is maintained throughout the life of this contract. 4. EQUIPMENT MAINTENANCE & RESPONSE TIMES The vendor shall provide full-service maintenance through a service agreement for minimum 2 years post installation. The vendor shall provide all additional recurring and preventive maintenance activities: Emergency response for equipment and MSPS failures. The vendor shall respond to the facility for any equipment failure resulting in a total or partial loss of the equipment installed within or any portion of the mobile sterile processing system within 4 hours from the time of VAMC contact. If a permanent solution cannot be resolved within 24 hours of initial failure, a temporary solution shall be operational within 24 hours of initial failure. An agreed permanent solution shall be planned, and replacements ordered within 5 business days of initial failure. 5. SAFETY The contractor/vendor shall always, ensure that all work provided complies with all local, State, and Federal rules and regulations pertaining to workplace safety. The contractor shall take all proper safety and health precautions to protect the work, the workers, the public and the property (OSHA). Installs and construction work will be in accordance with all applicable codes to include the most current codes National Fire Protection Association (NFPA), and the VA Design Guide Manuals that are applicable to the work. If the design guide is equivalent or more stringent then the mentioned codes, the VA Design Guide takes precedence. 6. DOCUMENTATION The contractor shall furnish to the KCVA Medical Center Biomedical Engineering department legible hard copies and electronic copies of service reports of the maintenance indicating details of the maintenance performed upon completion of all work. 7. PLACE OF PERFORMANCE Kansas City Veteran Administration Medical Center, 4801 Linwood Blvd. Kansas City 64128 8. PERFORMANCE MONITORING The contractor s performance will be monitored by the Contracting Officer (CO), Contracting Officer Representative (COR), Infection Control Risk Assessment Office, and Safety Office during the rental agreement period. Coordinate all administrative and construction requirements exclusively with or through assigned Contracting Officer Representative (COR). The Construction Safety Officer, Construction Safety Committee, and COR will review the operations and site construction conditions of the project to confirm that all Safety regulations relating to construction and healthcare facilities are being adhered to. Daily and weekly safety rounds will be conducted. For non-compliance safety related issues, a safety report will be sent to the Contractor to correct within 14 days. Non-compliance may result in job-site suspense until compliance is met at the contractor s expense. 9. HOURS OF OPERATION The contractor will perform services required during the hours of 7:30 am 4:00 pm, Monday through Friday or as directed by the designated COR. 10. SECURITY/ IDENTIFICATION The C&A requirements do not apply, and Security Accreditation Package is not required Taking of photographs/videos/recording of any kind is PROHIBITED on VA premises unless pertaining to work assignment, instructed or authorized by the Contracting Officer s Representative (COR).

From Solicitation posted on Apr 16, 2026

Notice history

1
  1. Solicitation LATEST Posted Apr 16, 2026 View

Details

Solicitation number 36C25526Q0323
Notice ID 45136410c38644f6b725679810f96c3d
Notice type Solicitation
Product / Service (PSC) 6515
NAICS 334510
Place of performance Kansas City, Missouri
Archive date May 13, 2026

Award Information

Not yet awarded

Documents

1
View on SAM.gov

Contacts

primary
Jon Lutmer

Email

Phone

Agency

VETERANS AFFAIRS, DEPARTMENT OF
VETERANS AFFAIRS, DEPARTMENT OF
255-NETWORK CONTRACT OFFICE 15 (36C255)

Place of Performance

Kansas City, Missouri 64128
USA

Dates

Posted Apr 16, 2026 3 months ago
Last Updated Aug 06, 2026 1 day ago
Due Apr 28, 2026 3 months ago